The method is based on “dilute and shoot” principle. Sample preparation involves: ⇰ The addition of the internal standard is performed in the needle of the autosampler, before injection into UHPLC-MS/MS.OBJECTIVE
This abstract presents a multi-mycotoxin sample preparation and analysis method for EU-regulated and emerging mycotoxins with Agilent 6460c UHPLC-MS/MS.MATERIALS & METHODS
Method validation in terms of linearity, selectivity, sensitivity, accuracy, and precision was performed in 8 complex feed matrixes (corn, compound feed, wheat, barley, soya meal, wheat bran, sunflower meal, total mixed ration) to detect 34 mycotoxins:
The limits of quantitation (LOQs) were from 0.05 ngmL-1 to 4 ngmL-1.
RESULTS & CONCLUSIONS
RESULTS & CONCLUSIONS
The recoveries of the 34 mycotoxins spiked in the 8 feed categories were in the range of 65.5%–125.7% after internal standard correction, with relative standard deviations (RSDs) below 15.1%.
Results showed the robustness of this method for analysis of 34 mycotoxins using LC-MS/MS.
Autores Hunor Farkaš, Svetlana Ćujić, Jog Raj, Zdenka Jakovčević, and Marko Vasiljević PATENT CO, DOO., Vlade Ćetkovića 1A, 24 211, Mišićevo, Serbia

45th Mycotoxin Workshop, June 2–5, 2024, Vienna, Austria